Customization: | Available |
---|---|
Customized: | Customized |
Scale: | Medium |
Still deciding? Get samples of US$ 4000/Set
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Digital Radiography portable100mA X Ray Machine with Dr System RD500A
Working conditions of the camera system:
1. Environmental conditions
(1) Operating temperature: 15ºC~35ºC
(2) Relative humidity: 30% to 75%, no condensation
2. Power conditions
(1) Power supply voltage: single-phase 220VAC±10%, 50Hz/60Hz
High voltage generator
1. Output power: 5.3KW
2. Tube voltage range: 40~125kV
3. Tube current range: 10mA ~ 100mA (using R'20 standard)
4. Loading time range: 1.0ms~4000ms (using R'20 standard)
5. Current time product range: 0.1 ~ 200mAs (using R'10 standard)
6. Maximum power: 5.3KW
7. Power supply: single-phase 220VAC±10%, 50Hz/60Hz
X-ray source components
1. Nominal tube voltage: 40KV~125KV
2. Focus: 0.6mm/1.8mm
3. Target angle: 15°
4. Anode heat capacity: 30KJ
Flat panel detector
1. Material: amorphous silicon
2. Scintillation body: Csl
3. A/D conversion: 16bit
4. Effective imaging area: 430mm x 430mm
5. Acquisition matrix: 3072 x 3072 pixels
6. Pixel size: 139μm
7. Drawing time: ≤3S
8. Spatial resolution: 3.6LP/mm
9. The weight of the detector: 3.75KG
10. Waterproof grade: IPX1
11. Working environment: 5-35ºC, 10-75%RH
12. Data interface: GigE