Digital X-ray Machines 2mm Lead Sheet

Product Details
Customization: Available
Classification: Imaging Diagnostic Equipment
Type: X Ray Equipment
Still deciding? Get samples of US$ 11000/Piece
Request Sample
Manufacturer/Factory, Trading Company, Other, Group Corporation

360° Virtual Tour

Diamond Member Since 2023

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Registered Capital
15000000 RMB
Plant Area
>2000 square meters
  • Digital X-ray Machines 2mm Lead Sheet
  • Digital X-ray Machines 2mm Lead Sheet
  • Digital X-ray Machines 2mm Lead Sheet
  • Digital X-ray Machines 2mm Lead Sheet
  • Digital X-ray Machines 2mm Lead Sheet
  • Digital X-ray Machines 2mm Lead Sheet
Find Similar Products
  • Overview
  • Product Description
  • Detailed Photos
  • Clinical Images
  • Company Profile
  • customers feedback
Overview

Basic Info.

Model NO.
RD-832X
Model No.
Rd-832X
Detector Size
17"*17"
Pixel
3072*3072
Output Power
32kw
Item
X Ray Machine
Exposure Kv
40kv-125kv
Tube Current Range
10mA ~ 400mA
Power Supply
100-240V, 50Hz/60Hz
Nominal Tube Voltage
125kv
Inverter Frequency
25kHz
Transport Package
Standard Export Package
Specification
3.5 M3
Trademark
Dawei
Origin
China
HS Code
9022140090
Production Capacity
50000

Product Description

 
Product Description

32kw 400mA Floor-Mounted Digital Radiography X-Ray machines
 

No. Component Performance parameter
1 Medical diagnostic X-ray high-frequency high-voltage generator Output Power: 32kW
Input Voltage: 220VAC,±10%,50Hz
Inverter Frequency: 25kHz,±10%
kV range: 40 kV~150kV
mA range: 10 mA~400 mA
ms range: 1ms~6300 ms
mAs range: 0.1mAs~150mAs
Tech mode: kV-mA(AEC);kV-mAs;kV-mA-ms
APR: 2000
AEC(Optional):  AEC Ionization chamber power supply:+/-15V
Minimum exposure time:5ms
Auxiliary power supply: +24V DC 2A;24V AC 8A
Communication Interface: RS-232
2 X-ray Tube Assembly X-ray Tube Voltage Range: 40kV~125kV
Nominal Focal Spot Value:
0.6mm(Small Focus)/1.2mm(Large Focus)
Nominal Anode Input Power:
20kW(Small Focus)/40kW(Large Focus)
Anode Heat Content: 150KHu
3 Flat Panel Detector Image Sensor: a-Si (Amorphous Silicon) TFT
Scintillator: CsI:Tl
Effective Array: 3072 ×3072
Pixel Size: 139 um
Effective Area (H x V): 427mm×427mm
Preview Time: 3s
Imaging Time: 5s
Spatial Resolution: 3.6 lp/mm
AD Conversion: 16 bit
Data Interface: Gigabit Ethernet
Trigger Mode: AED(optional)/Software/Prep
Dimension (L × W × H): 460mm×460mm×15mm
Weight: 4kg
Power Consumption: 20W
4 Collimator Control Mode: Manual control
Maximum Radiation: 150kV
Inherent Filtration Al. Equivalent: 1mm AL
Rectangular X-ray Field Selection 
Set Focus Distance, SID=100cm: Min 2cm×2cm;Max 43cm×43cm
X-Ray Field Deviation: ≤1% of SID
Light Field Indicator @ SID=100cm: ≥180lux
Light Field Indicator Precision: ≤1% of SID
Light source ON time: 30s
Maximum Radiation Leakage EN 60601-1-3: <0.5mGyh
Power Supply: 24V AC/DC,Max.4A
Dimension: 206mm×230mm×143mm
Weight: 7.0kg±0.5kg
5 Machine Wall Stand Motion control: Manual control
Overall height: 2125mm±10mm
BUCKY itinerary:
1450mm±10mm(The highest: 1625mm, the lowest: 335mm)
Starting force: ≤50N
Braking force: ≥100N
Tube Stand Motion control: Manual control
Overall height: 2125mm±10mm
Longitudinal stroke of tube column: 1280mm±10mm
Tube rotation range: ±180°
SID:1120mm±10mm
Starting force: ≤50N
Braking force: ≥100N
Radiographic Table Radiographic table height: 540mm±5mm
Table size (L×W): (2000mm±5mm)×(665mm±5mm)
BUCKY itinerary: Manual control, 1245mm±10mm
Table load-bearing: 200kg
6 Grid Grid 1 Grid density: 103 L/inch
Foucesing distance: 100cm
Grid ratio: 10:1
Grid 2 Grid density: 103 L/inch
Foucesing distance: 150cm
Grid ratio: 12:1
7 Workstation /
8 Dawei DR image acquisition and processing software This software is comprised of following modules which provides a work flow of patient study:
Patient Management: including patient registration, work list, study management.
Study operation: including bodypart selection, study items selection, image acquiring.
Image preview: including display, layout and processing of image. Also tool options for advanced operation.
Configuration: including configuration of system, study and user management. Especially the configuration for worklist and storage.  
The image processing tools are as follows:
Flip vertical, flip horizontal, rotate -90°, rotate 90°, free rotate, window/level change, auto WW/WL, ROI window/level, invert color, image stitching, reject image, zoom in/out, image advanced process, reset the image, invalid image, measure the angle, distance measure, point gray value of the image, ROI magnifier, magnifier, image hint, CTR(cardio-thoracic ratio) tool, arrow tool, copy tool, paste tool, email tool, cobb angle, monitor calibration tool, save as tool, compare tool and inverse compare tool.
Other tools: mark left, mark right, mark anterior, text edit, clip tool, clip options tool, ellipse clip tool and delete tool.
Detailed Photos

Digital X-ray Machines 2mm Lead SheetDigital X-ray Machines 2mm Lead Sheet

Digital X-ray Machines 2mm Lead SheetDigital X-ray Machines 2mm Lead SheetDigital X-ray Machines 2mm Lead Sheet

Clinical Images

 

Digital X-ray Machines 2mm Lead Sheet
Digital X-ray Machines 2mm Lead Sheet
Digital X-ray Machines 2mm Lead Sheet
 
Company Profile

Digital X-ray Machines 2mm Lead SheetDigital X-ray Machines 2mm Lead Sheet

customers feedback

 

Digital X-ray Machines 2mm Lead SheetDigital X-ray Machines 2mm Lead SheetDigital X-ray Machines 2mm Lead Sheet

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier