Digital X-Ray Radiography system RD-500B
No. | Component | Performance parameter | |
1 | Portable X-ray equipment | High-voltage Generator | Output power: 5.3kW |
Rated voltage: 220VAC ,±10%, 50Hz | |||
Inverter Frequency: 100kHz,±5% | |||
kV range: 40kV~125kV, 1kV step | |||
mA range: 10mA~100mA(R'20) | |||
ms range: 1ms ~4000ms(R'20) | |||
mAs range: 0.1mAs~200mAs(R'20) | |||
Exposure mode: kV-mAs;kV-mA-ms | |||
APR num:2000 | |||
Communication Interface: RS-232 | |||
Tube | Nominal Focal Spot Value: 0.6mm(Small Focus) 1.8mm(Large Focus) |
||
Exposure Duty-cycle: 1:60 | |||
Maximum X-ray Tube Current: 20mA(Small Focus) 100mA(Large Focus) |
|||
Radiation Protection: Leakage Technique Factor 125kV,0.24mA | |||
Construction: Tungsten | |||
Anode Angle: 15° | |||
Permanent Filtration: 0.6mmAl | |||
Additional Filtration for Portable Equipment: 1.5mmAl | |||
Collimator | Maximum X-ray field coverage range: 430mm×430mm(SID=90cm) | ||
The average brightness of the light field: >100lux | |||
Edge contrast ratio: >3:1 | |||
Lamp: LED lamp | |||
Bright X-ray field duration for Once: 30s | |||
Filtration(Inherent): 1mmAl | |||
Control method: Manual | |||
Light field size instructions: Pointer to the calibration | |||
overall unit | Volume: 480mm×320mm×223mm | ||
Weight: 17.4kg | |||
2 | Flat Panel Detector | Image Sensor: a-Si (Amorphous Silicon) TFT | |
Scintillator: CsI:Tl | |||
Effective Array: 3072 ×3072 | |||
Pixel Size: 139 um | |||
Effective Area (H x V): 427mm×427mm | |||
Preview Time: 3s | |||
Imaging Time: 5s | |||
Spatial Resolution: 3.6 lp/mm | |||
AD Conversion: 16 bit | |||
Data Interface: Gigabit Ethernet | |||
Trigger Mode: AED(optional)/Software/Prep | |||
Dimension (L × W × H): 460mm×460mm×15mm | |||
Weight: 4kg | |||
Power Consumption: 20W | |||
3 | Mobile machine | Machine Size (L×W): 1030mm×900mm,±5% | |
Machine Height Adjustment Range (Based on the height of the focus): 560mm~1590mm,±5% | |||
Machine Rotation Angle Range: 360° | |||
4 | Laptop | / | |
5 | Dawei DR image acquisition and processing software | This software is comprised of following modules which provides a work flow of patient study: Patient Management: including patient registration, work list, study management. Study operation: including bodypart selection, study items selection, image acquiring. Image preview: including display, layout and processing of image. Also tool options for advanced operation. Configuration: including configuration of system, study and user management. Especially the configuration for worklist and storage. The image processing tools are as follows: Flip vertical, flip horizontal, rotate -90°, rotate 90°, free rotate, window/level change, auto WW/WL, ROI window/level, invert color, image stitching, reject image, zoom in/out, image advanced process, reset the image, invalid image, measure the angle, distance measure, point gray value of the image, ROI magnifier, magnifier, image hint, CTR(cardio-thoracic ratio) tool, arrow tool, copy tool, paste tool, email tool, cobb angle, monitor calibration tool, save as tool, compare tool and inverse compare tool. Other tools: mark left, mark right, mark anterior, text edit, clip tool, clip options tool, ellipse clip tool and delete tool. |