Customization: | Available |
---|---|
Customized: | Customized |
Certification: | CE, ISO13485 |
Still deciding? Get samples of US$ 11000/Set
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Output Power: 32kW
Input Voltage: 220VAC,±10%,50Hz
Inverter Frequency: 25kHz,±10%
kV range: 40 KV~150KV
mA range: 10 mA~400 mA
ms range: 1ms~6300 ms
mAs range: 0.1mAs~150mAs
Tech mode: kV-mA(AEC);kV-mAs;kV-mA-ms
APR: 2000
AEC(Optional): AEC Ionization chamber power supply:+/-15V
Minimum exposure time:5ms
Auxiliary power supply: +24V DC 2A;24V AC 8A
Communication Interface: RS-232
|
|
X-ray Tube Voltage Range: 40kV~125kV
Nominal Focal Spot Value:
0.6mm(Small Focus)/1.2mm(Large Focus)
Nominal Anode Input Power:
20kW(Small Focus)/40kW(Large Focus) Anode Heat Content: 150KHu |
|
Image Sensor: a-Si (Amorphous Silicon) TFT
Scintillator: CsI:Tl
Effective Array: 3072 ×3072
Pixel Size: 139 um
Effective Area (H x V): 427mm×427mm
Preview Time: 3s
Imaging Time: 5s
Spatial Resolution: 3.6 lp/mm
AD Conversion: 16 bit
Data Interface: Gigabit Ethernet
Trigger Mode: AED(
optional)/Software/Prep
Dimension (L × W × H): 460mm×460mm×15mm
Weight: 4kg
Power Consumption: 20W
|
|
Control Mode: Manual control
Maximum Radiation: 150kV
Inherent Filtration Al. Equivalent: 1mm AL
Rectangular X-ray Field Selection
Set Focus Distance, SID=100cm: Min 2cm×2cm;Max 43cm×43cm
X-Ray Field Deviation: ≤1% of SID
Light Field Indicator @ SID=100cm: ≥180luxDigital X-ray Radiography System
3 / 5
Light Field Indicator Precision: ≤1% of SID
Light source ON time: 30s
Maximum Radiation Leakage EN 60601-1-3: <0.5mGyh
Power Supply: 24V AC/DC,Max.4A
Dimension: 206mm×230mm×143mm
Weight: 7.0kg±0.5kg
|
|
Motion control: Manual control
Overall height: 2125mm±10mm
BUCKY itinerary:
1450mm±10mm(The highest: 1625mm, the lowest: 335mm)
Starting force: ≤50N
Braking force: ≥100N
|
|
Motion control: Manual control
Overall height: 2125mm±10mm
Longitudinal stroke of tube column: 1280mm±10mm
Tube rotation range: ±180°
SID:1120mm±10mm
Starting force: ≤50N
Braking force: ≥100N
|
|
Radiographic table height: 540mm±5mm
Table size (L×W): (2000mm±5mm)×(665mm±5mm)
BUCKY itinerary: Manual control, 1245mm±10mm
Table load-bearing: ≥135kg
|
|
Grid density: 103 L/inch
Foucesing distance: 100cm
Grid ratio: 10:1
|